  • BD Vacutainer Blood Collection Tubes Class I Recall

    Tuesday, March 27, 2018

    BD is recalling their Vacutainer EDTA Blood Collection Tubes with lavender, tan, pink and green rubber tube stoppers due to a chemical in the rubber tube stopper that interferes with the accuracy of the Anodic Stripping Voltammetry (ASV) testing methodology. ASV is the methodology used in Magellan Diagnostics’ LeadCare Testing Systems. The tube stoppers contain a substance called thiuram that can sometimes release sulfur-containing gases, which may dissolve into the blood sample and bind the lead particles. This chemical reaction makes it difficult for the Magellan lead tests to detect the correct amount of lead in the sample and may cause falsely lower test results. Falsely lower test results may lead to improper patient management and treatment for lead exposure or poisoning. The use of affected product may cause serious adverse health consequences.

    Read More
  • Methylprednisolone Sodium Succinate for Injection 40mg, 125mg, and 1g by Sagent Pharmaceuticals: Recall

    Wednesday, March 7, 2018

     

    Sagent Pharmaceuticals, Inc. announced the voluntary nationwide recall of ten lots of Methylprednisolone Sodium Succinate for Injection, USP, 40mg, 125mg, and 1g. A detailed listing of products and lots is listed in the recall notice. These products were manufactured by Gland Pharma Ltd. and distributed by Sagent Pharmaceuticals. Sagent has initiated this voluntary recall of Methylprednisolone Sodium Succinate for Injection, USP to the user level due to the discovery of high out of specification impurity results detected during routine quality testing of stability samples for two lots. 